|
Model Number 317.82 |
Device Problems
Break (1069); Entrapment of Device (1212)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/13/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional narrative: additional device product codes: erl and hbe.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, when the surgeon pierced the patient's bone, the drill broke, leaving the tip inside the patient's bone.This report involves one (1) 1.5mm drill bit/mini qc with 12mm stop/44.5mm.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: part: 317.820 lot: f-29924 manufacturing site: selzach supplier: sphinx werkzeuge ag release to warehouse date: march 23, 2020 a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the 1.5mm drill bit/mini qc with 12mm stop/44.5mm was returned and received at us customer quality (cq).Upon visual inspection, the drill bit tip was broken, and the broken pieces were not returned.However, the embedded condition cannot be confirmed since no x-rays were provided.No other issues were identified with the returned components of the device.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Document/specification review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed.No design issues or discrepancies were identified.Investigation conclusion the complaint condition is confirmed for the 1.5mm drill bit/mini qc with 12mm stop/44.5mm.While a definitive root cause could not be identified for the reported problem, it is possible that the device might have encountered unintended forces.There is no indication that a design or manufacturing issue has caused the complaint condition based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- product complaint # (b)(4).Investigation summary =product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot = part: 317.820 lot: f-29924, manufacturing site: selzach, supplier: (b)(4), release to warehouse date: 23.Mar.2020.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|