Model Number N/A
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event was initially reported on 0001822565-2020-02530 and 0001825034-2020-04179.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows the drill pin stuck in a hole.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the drill pin got stuck to the device during an initial procedure.Attempt for further information has been made, but no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL CUT GUIDE
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11678493
MDR Text Key245819096
Report Number0001822565-2021-01039
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00889024582859
UDI-Public(01)00889024582859(10)64768305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200151
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42539905295
Device Lot Number64768305
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-