MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problem Difficult or Delayed Positioning (1157)
Patient Problems Cardiac Arrest (1762); Low Blood Pressure/ Hypotension (1914); Perforation (2001); Nervous System Injury (2689); Pericardial Effusion (3271)
Event Date 03/23/2021
Event Type  Death  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional clips referenced are filed under separate medwatch report number(s).
 
Event Description
This is filed to report pericardial effusion, perforation requiring open heart surgery.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with grade 4+ and mean pressure gradient of 6 mmhg.The patient has severe mitral regurgitation (mr) and was turned down by two sites for heart transplant; therefore, it was decided to perform the mitraclip procedure to stabilize the patient for heart transplant.Prior to the procedure, a small pericardial effusion was noted but it was very small and was not addressed.The clip delivery system (cds) (01201u103) was advanced; however, an aorta hugger was noted due to the patient anatomy and there was no support from the septum.Troubleshooting was performed and the cds was positioned.Grasping was performed with difficulty, and the clip got stuck in the chords.The clip was not able to be completely freed and was deployed on the posterior leaflet and chordae.There was no change to mr.It was decided to remove the steerable guide catheter (sgc) (10108u234) and the cds and make a new transseptal puncture that would provide more support.The new transseptal puncture was made in a thicker section of the septum which had more height and support.The same sgc was advanced without issue and the second cds (01201u104) was advanced to the valve.The clip was deployed, reducing mr to 2+.When the sgc was removed, the patient blood pressure dropped and the patient flatlined, cardiopulmonary resuscitation (cpr) was performed.Echocardiogram showed the pericardial effusion had gotten bigger and it was unknown where it was coming from.After 15 minutes of cpr and additional medication, the patient became stable and the patient remained stable for 45 minutes.It was then noted that there was a small hole where the second transseptal puncture had been made.Therefore, the team decided to treat the patient with open heart surgery and stitch the small hole in the septum.The patient was placed on extracorporeal membrane oxygenation (ecmo) and the chest was closed.The patient remains stable, two clips implanted reducing mr to 2+ and mean pressure gradient of 6 mmhg.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other incidents reported from this lot.Based on the available information, the reported difficult or delayed positioning with the anatomy appears to have been related to patient anatomy.The reported perforation appears to have been related to procedural conditions.The reported pericardial effusion, hypotension, and cardiac arrest appear to have been cascading events of the reported perforation.A cause for the reported nervous system injury and death could not be determined.The reported patient effects of perforation, pericardial effusion, hypotension, cardiac arrest, and death, as listed in the mitraclip system instructions for use, are known possible complications associated with mitraclip procedures.The reported hospitalization or prolonged hospitalization, unexpected medical interventions, required medication, and surgical intervention were results of case-specific circumstances.This event was further reviewed by an abbott medical affairs director ((b)(6)).The reviewer stated that: ¿there is no evidence that death was related to the device but was likely related to the procedure since the need for surgery was prompted by the pericardial effusion." there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
Subsequent to the initial report filing, additional information was received reporting that on (b)(6) 2021 the patient died due to no brain activity.The physician suspected that the pericardial effusion became bigger.The clips remained stable and attached to the leaflets.The patient had no brain activity and the family decided to withdraw care and the patient passed away.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11679194
MDR Text Key245811500
Report Number2024168-2021-03209
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/07/2022
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number10108U234
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2021
Initial Date FDA Received04/16/2021
Supplement Dates Manufacturer Received04/20/2021
Supplement Dates FDA Received05/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1 IMPLANTED MITRACLIP
Patient Outcome(s) Death; Hospitalization; Required Intervention; Disability;
Patient Age7 YR
Patient Weight51
-
-