Catalog Number R5C4479 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the cassette of a homechoice automated pd set with cassette was cracked.The customer further described the event as, ¿unable to inject solution due to the rectangular plastic piece cracked¿.This was observed prior to use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection with naked eye noted a portion of the welded cassette with open seal.An underwater pressure test was performed and a leak was observed from the open seal location.The reported condition was verified.The cause of the condition was related to manufacturing.Should additional relevant information become available, a supplemental report will be submitted.
|
|