• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 STATLOCK PICC PLUS CRESCENT SLIDING POST; STAND, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 STATLOCK PICC PLUS CRESCENT SLIDING POST; STAND, INFUSION Back to Search Results
Catalog Number PIC0220
Device Problems Misconnection (1399); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of juey1044 showed one other similar product complaint(s) from this lot number.The complaints for this lot number have been reported from the same facility in (b)(6).
 
Event Description
It was reported that one wing of the statlock device does not stay closed.No other information was provided.
 
Event Description
It was reported that one wing of the statlock device does not stay closed.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint that one wing of the statlock stabilization device would not stay closed was confirmed, but the exact cause could not be determined.One statlock was received in open packaging.The adhesive liner had not been removed from the substrate.One side of the retainer was closed.The door on the other side of the retainer was open and one securement latch was bent outwards.The hinge exhibited a crease, which indicates that the door had been moved from its open position.This type of damage can occur if the doors are misaligned when locking the retainer; however, the exact cause could not be determined.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK PICC PLUS CRESCENT SLIDING POST
Type of Device
STAND, INFUSION
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11682736
MDR Text Key246286281
Report Number3006260740-2021-01407
Device Sequence Number1
Product Code FOX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPIC0220
Device Lot NumberJUEY1044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Initial Date Manufacturer Received 03/26/2021
Initial Date FDA Received04/16/2021
Supplement Dates Manufacturer Received06/02/2021
Supplement Dates FDA Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-