Catalog Number 66800472 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, foot pedal was not recognized by versajet ii power console.A replacement foot pedal was tested and functioned properly.No case involved; therefore, no patient involvement.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.Probable causes may include a component failure.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found further instances of the reported events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
|