Code orange called over head for chemical spill/leak in medical procedure unit (mpu) scope room.Biomed technician called as well.Found chemical was coming from underneath the scope disinfector (dsd).Biomed technician ran a simulated scope run to find leak.Found a split in the internal tubing line on the back of the chemical tray.This line would leak when trying to pump disinfection to the basin and also drain the basin when the pump stopped.This is where the fluid on the ground came from.Biomed examined the tubing lines of our second dsd machine.Found these lines had deterioration as well.These lines are not called out to be replaced in the preventive maintenance instructions.We have replaced all tubing lines in both machines.
|