Catalog Number R5C4479
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The actual device was not available; however, three (3) photographs of the sample were provided for evaluation.Visual inspection of the photographs noted a crack at the side of the welded cassette.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice cassette had a "crack on the side of the cassette".There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key11685399
MDR Text Key246019953
Report Number1416980-2021-02205
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412030531
UDI-Public(01)00085412030531
Combination Product (y/n)Y
Reporter Country CodeDA
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479
Device Lot NumberS20E22093
Was Device Available for Evaluation? No
Date Manufacturer Received03/25/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-