It was reported that during a cryoablation procedure, the patient had a sudden onset of 'aspen syndrome,' upper body twitching, and confusion.The case was aborted.It was noted the patient's heartbeats were weak.Ultrasound indicated the presence of a pericardial effusion.A pericardial tamponade was noted.A pericardiocentesis was performed.The patient had cardiac arrest.Cardiopulmonary resuscitation was performed.A blood transfusion was given.The patient's vital signs stabilized under observation, and the patient was transferred to the intensive care unit (icu).No further patient complications have been reported as a result of this event.
|
Product event summary: the sheath 4fc12 with lot number 0010305410 was returned and analyzed.Visual inspection of the sheath showed the device shaft was intact with no apparent issue.Using the sentinel blackbelt leak tester, the sheath passed the test.The shaft and valve were leak tight with no apparent issues.The aspiration and flushing test did not show any air passing through the tube or expelled from the sheath distal tip; the hemostatic valve was leak tight.Deflection worked as per specification.Several clinical issues (pericardial effusion, cardiac tamponade, cardiac arrest or cardiopulmonary arrest, neurological deficit/function loss, confusion ) occurred during the procedure.In conclusion, the reported clinical issues cannot be assessed through testing.There is no indication of a relation of the adverse events to the performance and malfunction of the product.The sheath passed the returned product inspection as per specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|