Model Number APC 2 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
Bowel Perforation (2668)
|
Event Date 03/26/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The apc/esu system was returned and thoroughly inspected/tested.The apc/esu system was returned and thoroughly inspected/tested.A review of the chronological data at the time of the event didn't show that there were any equipment problems during the procedure.Then, a technical safety check was performed on each unit.This included an electrical safety check, a function check of each of the equipment's features, and a power output check.Also, the gas flow rates were measured and found to be within their acceptable ranges for the apc.All features were/are functioning properly within specifications on both devices [note: unrelated to the reported issue, the esu was updated to the software version (b)(4).This is typical routine service work.].In addition, no anomalies were found in the device history records (dhrs) for the apc and esu.In conclusion, no erbe equipment problem was found that would have caused or contributed to the event.Most likely, there were many factors involved in the reported incident.Specifically, as reported the patient's condition of being high risk for the interventional work was a significant factor in the event.Nevertheless, upon the intervention, the remaining tissue of the bowl did not stay intact which resulted in the perforation.The account is being made aware of the findings.To further address the issue, additional in-service work was offered to the medical staff at the hospital; however, it was declined as they are very experienced users.No trends have been identified.Erbe usa, inc.Is now closing the file on this event.
|
|
Event Description
|
It was reported that an erbe system, argon plasma coagulator (apc)/electrosurgical unit (esu/generator, model vio 300 d, part number (p/n) 10140-100, serial number (b)(4)) system was involved in a patient incident upon a colonoscopy.The apc/esu system was used with a filter integrated argon plasma coagulation (fiapc) probe.The equipment settings were pusled apc mode, effect 2, 20 watts, 0.5 liters per minute (lpm) flow rate.Per the account, the patient was high risk for the interventional work.Specifically, patient had multiple comorbidities like strictures, being very elderly, etc.Upon the interventional work a perforation occurred.To address the issue, surgery was performed to close the perforation.
|
|
Search Alerts/Recalls
|
|