• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO UKR X3 ONLAY INSERT SIZE 5 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO UKR X3 ONLAY INSERT SIZE 5 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180735-1-E
Device Problem Insufficient Information (3190)
Patient Problem Unspecified Infection (1930)
Event Date 03/24/2021
Event Type  Injury  
Manufacturer Narrative
Reported event: an event regarding infection involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.There have been no other similar events for the sterile lot referenced.Conclusions: all stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance to applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.The following devices were also listed in this report: 180514;mck femoral-rm-ll-sz 4;2738-1.180615;mck tibial baseplate-rm/ll-sz 5;126140220-01.6194-1-001;simplex hv us 1 pack;1013ab843cb.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event Description
Dr.Performed an i&d poly exchange on a right mako medial uni originally done (b)(6) 2021 due to infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO UKR X3 ONLAY INSERT SIZE 5 - 8 MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
joanne mahony
ida industrial estate
carrigtwohill NA
EI   NA
214532800
MDR Report Key11694932
MDR Text Key246810766
Report Number3005985723-2021-00072
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier07613327386028
UDI-Public07613327386028
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150307
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number180735-1-E
Device Catalogue Number180735-1-E
Device Lot NumberEN747D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2021
Initial Date FDA Received04/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
-
-