The reported complaint of "the autopulse platform (sn (b)(4) ) intermittently powered off was not confirmed during the archived data review or during the initial functional testing.The autopulse platform passed the functional testing and the autopulse platform worked as intended.Visual inspection was performed and observed pieces of white plastic foam in the battery bay that could be a possible cause of the platform's intermittent power loss.In addition, the power button of the platform had evidence of liquid ingress that could also be a possible cause of the power loss.This type of observed issue is likely attributed to mishandling.The foam was removed from the battery bay and as a precautionary measure, the power button and cable were replaced.Also, unrelated to the reported complaint, found a damaged front enclosure as a result of user mishandling such as a drop.The front enclosure was replaced to address the damage.The platform passed the initial functional testing without any fault or error.The archive data review showed no significant discrepancies.Following service, the autopulse platform passed the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The autopulse platform passed the final testing without any fault or error.
|
During the shift check, the autopulse platform (sn (b)(4) ) intermittently powered off without displaying any user advisories.The user tried multiple batteries, however, the issue persists.No patient involvement.
|