Model Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the companion 2 driver did not pass the system check.
|
|
Manufacturer Narrative
|
The customer-reported issue was confirmed by review of the patient data file.The system checkout logs revealed three system check test failures which reported a 9-volt battery reading below acceptable limits (6.500 v - 10.000 v).During investigation testing, the driver did not pass the system check for this same reason thus replicating the customer-reported issue.The 9-volt battery is replaced during servicing after 90 days of driver use or at the two-year service interval regardless of days of use.The 9-volt battery was replaced during the most recent service completed on 06 aug 2019.The 9-volt battery installed at the time of service had an expiration date of march 2021.The driver was shipped on 14 oct 2019 and was not returned for service prior to the battery's expiration date because the driver had not accumulated 90 days of patient support.It is suspected that the 9-volt battery experienced normal depletion below the specified 6.5 volts.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5507 follow-up report 1.
|
|