MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2021
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the companion 2 driver did not pass the system check.
 
Manufacturer Narrative
The customer-reported issue was confirmed by review of the patient data file.The system checkout logs revealed three system check test failures which reported a 9-volt battery reading below acceptable limits (6.500 v - 10.000 v).During investigation testing, the driver did not pass the system check for this same reason thus replicating the customer-reported issue.The 9-volt battery is replaced during servicing after 90 days of driver use or at the two-year service interval regardless of days of use.The 9-volt battery was replaced during the most recent service completed on 06 aug 2019.The 9-volt battery installed at the time of service had an expiration date of march 2021.The driver was shipped on 14 oct 2019 and was not returned for service prior to the battery's expiration date because the driver had not accumulated 90 days of patient support.It is suspected that the 9-volt battery experienced normal depletion below the specified 6.5 volts.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5507 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key11701327
MDR Text Key246507620
Report Number3003761017-2021-00046
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Date Manufacturer Received04/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-