The bwi product analysis lab received the device for evaluation on 1/28/2021.The device evaluation was completed on 4/7/2021.The product was returned to biosense webster for evaluation.Bwi conducted a visual inspection and force reading evaluation of the returned device.Visual analysis of the returned sample revealed reddish material in the pebax was observed on the smart touch bidirectional sf catheter.Magnetic sensor functionality was tested on carto and the catheter was properly visualized and no errors were observed.Then, the force sensor feature was tested and it was working properly, the force values were observed within specifications.Sem analysis show evidence of mechanical damage and a hole on the pebax surface.Object that cause the damage is unknown.No other anomalies were observed.A manufacturing record evaluation was performed, and no internal actions related to the reported complaint condition were identified.As part of bwi¿s quality process all devices are manufactured, inspected, and released to approved specifications.The instructions for use contain the following warning stated in the carto 3 system manual: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.Regarding the additional finding observed, the instruction for use contains the following information: in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
On january 28, 2021, this thermocool® smart touch® sf bi-directional navigation catheter was received by the biosense webster product analysis lab as an extra product under manufacturer reference number (b)(4).Manufacturer reference number (b)(4) was assessed as not reportable for a force issue.The visual finding on (b)(6) 2021 from the biosense webster failure analysis lab for this extra product received was that there was reddish material found in the pebax.Therefore, additional analysis was performed.The scanning electron microscope (sem) analysis showed evidence of mechanical damage and a hole on the pebax surface.Object that caused the damage was unknown.No other anomalies were observed.This hole on the pebax was assessed as a reportable lab finding.An investigation was performed to determine if there was an existing manufacturer reference number for this extra product received.Additional information was received stating that this product did not belong to (b)(4) complaint.In addition, we were unable to determine if there was an existing manufacturer reference number for this device.Since the returned catheter condition has been assessed as a reportable malfunction, we have made the decision to create a related manufacturer reference number under (b)(4) to conservatively report this returned catheter condition.The awareness date for this reportable lab finding is march 26, 2021.
|