Model Number 66801280
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during npwt, a renasys tch pump was not charging when plugged in.The power cord was tapped to the device to provide intermittent charge while the patient walked around house, however it is claimed that a blue flame and sparking was visible at the power port connection.This generated damage at the power cord tip and the power input of the device.The device was replaced with a new pump to finish the treatment.No patient injury or other complications were reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.Probable root cause maybe a component failure.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TCH DEVICE&POWER SPLY
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key11712908
MDR Text Key246960374
Report Number8043484-2021-00928
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00040565126586
UDI-Public00040565126586
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66801280
Device Catalogue Number66801280
Date Manufacturer Received07/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-