It was reported that during npwt, a renasys tch pump was not charging when plugged in.The power cord was tapped to the device to provide intermittent charge while the patient walked around house, however it is claimed that a blue flame and sparking was visible at the power port connection.This generated damage at the power cord tip and the power input of the device.The device was replaced with a new pump to finish the treatment.No patient injury or other complications were reported.
|
H3, h6: the device, used in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.Probable root cause maybe a component failure.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|