• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH45
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was identified.The blade was fractured and separated from the device tip.The teflon pad, handle and contact rings appear to be intact.The jaw crimp was inspected prior to disassembly and was found to be bent and contacting the blade.The device was disassembled to inspect the scalpel rod.Gouges were identified on the blade outside of the cutting area as a result of contact with the shaft assembly.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the reported event could be attributed to: applying improper or unnecessary directional or rotational force to connect instrument to hand piece failure mode: jaws/blade subassembly damage or separation too much force or torque applied to instrument, or grasping/pulling the instructions for use(ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post market surveillance.
 
Event Description
It was reported that a piece fell off inside the patient.The piece was retrieved without needing an x-ray to locate it or needing a larger incision to retrieve it.There was no patient injury or medical intervention and extended procedure time reported was less than five minutes to replace the device.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key11713696
MDR Text Key265167897
Report Number0001056128-2021-00020
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346244
UDI-Public07613327346244
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161693
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2023
Device Model NumberHARH45
Device Catalogue NumberHARH45RR
Device Lot Number12429860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-