Model Number 227214
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that wireless pedals loosing connection.Per service manual operational and diagnostic, this complaint can be confirmed.It was found during evaluation that the transmitter pcb was defective, and footswitch failed pairing.Further, battery tray assembly was corroded, and bump on pads were missing.Additionally, minor scratches were identified with the device upon decontamination.The transmitter pcb, battery tray assembly and bump on pads were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Fluid ingress into the device and contact with the battery tray assembly is responsible for the corrosion.User mishandling might be the most probable root cause for the missing bump on pad and the minor scratches on device.The corroded battery tray assembly and defective transmitter pcb might be the most probable root cause for the pairing failure of the foot switch.There are no indications from this complaint investigation that the issue/ failure is manufacturing-related, therefore a manufacturing record evaluation is not required.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the sales rep that during a shoulder procedure on (b)(6) 2021, it was observed that the foot switch device was losing connection.During in-house engineering evaluation, it was determined that the battery tray assembly was corroded.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR VUE WIRELESS FOOTSWITCH
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11716808
MDR Text Key261683212
Report Number1221934-2021-01299
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705009657
UDI-Public10886705009657
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number227214
Device Catalogue Number227214
Device Lot Number1606138
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2021
Date Manufacturer Received03/31/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial