HOLOGIC, INC. MYOSURE TISSUE REMOVAL DEVICE; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 30-401LITE
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that during a myosure procedure, the physician observed that the blade of the myosure ¨broke off from the device¨.To be sure that any pieces didn´t fell inside the uterine cavity the physician performed additional hysteroscopy to look for the blade or pieces but couldn´t find any debris inside the uterus.No other information is available.
 
Manufacturer Narrative
The device was received and tested.It was observed during visual inspection that the cutting window was open and foreign material along with some markings were present in the inner surface of the outer tube.Once the device was connected to the control unit, the cutting window closed and the device performed as intended.The device was dissected to analyze the integrity of the blade and no additional findings were made.The blade was not damaged nor broken.As a result, the reported observation could not be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE TISSUE REMOVAL DEVICE
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
MDR Report Key11719691
MDR Text Key249035105
Report Number1222780-2021-00078
Device Sequence Number1
Product Code HIH
UDI-Device Identifier15420045505070
UDI-Public(01)15420045505070(10)20M04R(17)231204
Combination Product (y/n)N
PMA/PMN Number
K142029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/04/2023
Device Model Number30-401LITE
Device Catalogue Number30-401LITE
Device Lot Number20M04R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Date Manufacturer Received04/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-