Model Number 30-401LITE |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
|
Event Description
|
It was reported that during a myosure procedure, the physician observed that the blade of the myosure ¨broke off from the device¨.To be sure that any pieces didn´t fell inside the uterine cavity the physician performed additional hysteroscopy to look for the blade or pieces but couldn´t find any debris inside the uterus.No other information is available.
|
|
Manufacturer Narrative
|
The device was received and tested.It was observed during visual inspection that the cutting window was open and foreign material along with some markings were present in the inner surface of the outer tube.Once the device was connected to the control unit, the cutting window closed and the device performed as intended.The device was dissected to analyze the integrity of the blade and no additional findings were made.The blade was not damaged nor broken.As a result, the reported observation could not be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.
|
|
Search Alerts/Recalls
|