Model Number 1400
Device Problem Infusion or Flow Problem (2964)
Patient Problem Insufficient Information (4580)
Event Date 03/23/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: one smiths medical cadd-legacy duodopa ambulatory infusion pump was returned for analysis in used condition.Visual inspection showed front and rear housing were scratched and the lock was worn.Delivery accuracy tests were performed and the pump was found to be delivering properly and within specification.The pump passed all tests and was found to be operating properly.Based on the investigation, the complaint allegation was not confirmed.
 
Event Description
Information was received indicating that a smiths medical cadd-legacy duodopa ambulatory infusion pump may have been delivering inaccurately.Per reporter some cassettes were "almost half and others are almost empty." it was reported that there was no variation in the administered extra doses.Pump was subsequently exchanged.Programmed settings on pump reported as: morning dose 17ml, continuous dose 3.6ml, extra dose 3.5ml.Patient reported to take mirapexim for breakfast and dinner, madopar 0.5 before bedtime and in the morning.Extra dose administered at 19:00 hours per neurologist's indication.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11720817
MDR Text Key247208879
Report Number3012307300-2021-03428
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1400
Device Catalogue Number21-1400-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/26/2021
Initial Date FDA Received04/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age74 YR
Patient Weight57
-
-