Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Date 11/27/2020
Event Type  Injury  
Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event Description
The following article was reviewed by gore: "renal stent complications and impact on renal function after standard fenestrated endovascular aneurysm repair." authors: celine deslarzes-dubuis, kenneth tran, benjamin d.Colvard, jason t.Lee, annals of vascular surgery, volume 72, 2021, pages 106-113.A 60 total of gore® viabahn® vbx balloon expandable endoprosthesis (vbx) and 133 icast devices were utilized as bridging stents.17 total renal stent-related complications were reported including stent occlusions (n=4), stenosis (n=9), dislocation (n=3), and endoleak (n=1).All stent complications underwent reintervention with a technical success of 94.2%.Patients with occlusion were treated the day of diagnosis with a mean time from diagnosis to reintervention of 21.5 days.The study concludes with stating that: "fevar is a durable option for the treatment of juxtarenal aortic aneurysms and is associated with excellent secondary patency.Renal stent complications have no significant impact on renal function, but smaller native renal arteries are at higher risk of stent-graft complications." this report is covering the allegations of migration (3) with reported reintervention in the renal arteries of multiple patients (none of which have specifically identified a vbx device as the medical device associated with the allegation).Further information has been requested but has not been made available at this time.
 
Manufacturer Narrative
Updated to include the following additional information: patient information a2 and a3 fields.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Type of Device
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
MDR Report Key11721008
MDR Text Key249102151
Report Number2017233-2021-01897
Device Sequence Number1
Product Code PRL
Combination Product (y/n)Y
PMA/PMN Number
P160021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,litera
Type of Report Initial,Followup,Followup
Report Date 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/26/2021
Initial Date FDA Received04/23/2021
Supplement Dates Manufacturer Received03/26/2021
03/26/2021
Supplement Dates FDA Received07/21/2021
10/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age74 YR
-
-