BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Device Problems Mechanics Altered (2984); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Event Description
It was reported that, during navio preventative maintenance, they got a intermittent reading and null reading, and the handpiece test would end without completing the full motor stepper test.There was no patient involvement.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio handpiece part number 110137, sn (b)(6) intended for treatment was returned for evaluation.A relationship between the reported event and the device was confirmed.The reported problem was visually confirmed.The snap lock nut is broken at the threads.Although the reported problem was visually confirmed, a functional evaluation could not be performed due to the snap lock nut is broken.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is a mechanical failure of the snap lock nut.As a part of corrective action, the old style snap lock is being replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11735131
MDR Text Key248363431
Report Number3010266064-2021-00308
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
K152574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2021
Initial Date Manufacturer Received 04/06/2021
Initial Date FDA Received04/27/2021
Supplement Dates Manufacturer Received08/04/2021
Supplement Dates FDA Received08/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-