Model Number 71434403
Device Problems Naturally Worn (2988); Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Event Description
It was reported that during a tka the lgn distal fem cutting block was worn out and did not stay tightened down.The procedure was successfully completed without delay using the same device.This happened when the instruments were out of the patient hence no patient injury or other complications were reported.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex screw slot of the device's knob is severely worn preventing the device from functioning as intended.The device also has several nicks and scratches as well as wear.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN DISTAL FEM CUTTING BLOCK
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED.
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11735232
MDR Text Key247685872
Report Number1020279-2021-03546
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010550163
UDI-Public03596010550163
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71434403
Device Catalogue Number71434403
Device Lot Number12CM00399
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Initial Date Manufacturer Received 04/06/2021
Initial Date FDA Received04/27/2021
Supplement Dates Manufacturer Received05/18/2021
Supplement Dates FDA Received05/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-