Model Number CI-1601-05 |
Device Problems
Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Failure of Implant (1924); Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
|
Event Date 03/24/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing pain over the implant site and retention issues due to swelling at the implant site.From (b)(6) 2021, to (b)(6) 2021, the recipient was given a 10 day course of antibiotics, however, the issue did not resolve.External equipment was exchanged, however, the issue did not resolve.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The recipient reportedly has an infection.Revision surgery will be scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient will be reimplanted at a later date.The recipient has healed and the infection resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts on the top cover and bottom cover, the fantail, and electrode.In addition, the electrode was severed prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|