Model Number CI-1601-05
Device Problems Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
Event Date 03/24/2021
Event Type  Injury  
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event Description
The recipient is reportedly experiencing pain over the implant site and retention issues due to swelling at the implant site.From (b)(6) 2021, to (b)(6) 2021, the recipient was given a 10 day course of antibiotics, however, the issue did not resolve.External equipment was exchanged, however, the issue did not resolve.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
The recipient reportedly has an infection.Revision surgery will be scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient will be reimplanted at a later date.The recipient has healed and the infection resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts on the top cover and bottom cover, the fantail, and electrode.In addition, the electrode was severed prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key11742407
MDR Text Key247904840
Report Number3006556115-2021-00500
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862267
UDI-Public(01)07630016862267(11)190916(17)220831
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2022
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age11 YR
-
-