A physician reported that the perforator failed to disengage during cranial surgery resulting in dura mater damage.Hemostasis was performed.The procedure was completed with the device.The manufacturer of the drill used with the perforator was unknown.It is unknown if the drill was electric or pneumatic, it is unknown if the perforator clicked into place in the drill, and it is unknown if the recommended spring tests were being performed between each burr hole.No further information was provided by hospital.
|
The perforator was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.Failure analysis - the perforator unit was inspected using the unaided eye.The unit was heavily soiled with organic matter, no other anomalies were observed."ifu" testing procedure was performed.The hudson end and the inner/outer portions were fused, but freed.The unit performed as intended once the unit was freed.Testing included: applying adequate pressure on the perforator point, ensuring engagement occurs as the hudson end is rotated.When engagement occurs, placing thumb pressure on the perforator point to ensure a smooth, positive spring action.Ensuring the hudson end rotates smoothly within the perforator body when the unit is in the disengaged position.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.In the failure analysis that was performed, the returned unit was found to work as intended, and met all acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.
|