• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. BIOMET BONE CEMENT R 40 -3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. BIOMET BONE CEMENT R 40 -3 Back to Search Results
Catalog Number 3035890011-3
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).Pictures have been received.The reported event was confirmed as it can be seen that the cement powder leaked outside the inner pouch.The product analysis can't be performed as the product was not returned.The complaint is related to a well-known event, previously evaluated and investigated.No further investigation is required as per 21 cfr part 820.198 complaint files §(b) and (c).Investigation results concluded that the reported event was due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.A summary of the investigation was sent to the complainant conveying zimmer biomet conclusions.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that, before the surgery , the inner package was leaking no adverse event has been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET BONE CEMENT R 40 -3
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11751646
MDR Text Key250449742
Report Number3006946279-2021-00075
Device Sequence Number1
Product Code LOD
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number3035890011-3
Device Lot NumberA848DH1703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-