Model Number 1458Q/86 |
Device Problems
Difficult to Insert (1316); Use of Device Problem (1670); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/16/2021 |
Event Type
malfunction
|
Event Description
|
During the initial implant procedure, low pacing impedance was noted on the left ventricular (lv) lead.The physician also noted the abnormal curvature of the patient¿s anatomy made placement of the lead difficult.The lv lead was explanted and a new lead was implanted to resolve the event.The patient was stable.
|
|
Event Description
|
Additional information was received that the left ventricular (lv) lead had difficulty advancing with the guidewire.
|
|
Manufacturer Narrative
|
The reported events were low pacing impedance and guidewire insertion difficulty.As received, a complete lead without the guidewire was returned in one piece for analysis.The reported event of low pacing impedance was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.The reported event of guidewire insertion difficulty was confirmed.X-ray inspection of the lead found the damaged coil at the s-curve region.The cause of the reported event guidewire insertion difficulty was isolated to the damaged coil found on the lead consistent with procedural damage.
|
|
Search Alerts/Recalls
|