Model Number PFSR110137
Device Problem Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2020
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that before a tka procedure, when performing the handpiece test, the torque was in the 55 range and wouldn¿t do the faster speed tests.It was swapped for its backup to continue with the procedure without delays.No other complications were reported.The investigation found that the issue was due to a faulty motor (weak and worn out).
 
Manufacturer Narrative
H6: the navio handpiece p/n 110137 s/n (b)(6) intended for use in treatment was returned for evaluation.Nothing was identified visually that contributed to the reported complaint.A handpiece test was performed and failed.The reported problem was confirmed.A handpiece thermal test was performed and the handpiece motor seized at various points during the test.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is a mechanical component failure with the motor.Further investigation into the reported failure is being conducting to determine if additional actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11752240
MDR Text Key248363760
Report Number3010266064-2021-00320
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Date Manufacturer Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-