Model Number CD3267-40Q |
Device Problem
Delayed Charge Time (2586)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.There was no battery performance alert that was triggered.Estimated battery longevity was 5 months remaining.Device having been implanted since (b)(6) 2013 for ~7.5 years.
|
|
Event Description
|
It was reported that although previous reports of the battery appeared to be tracking normally, the charge time had been extending with each transmission over the past two years.The device was pending explant for prolonged charge time.The patient was stable.
|
|
Event Description
|
New information received notes the device was explanted and replaced.The patient was stable throughout.
|
|