MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA1D1
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problems Arrhythmia (1721); Device Overstimulation of Tissue (1991)
Event Date 07/31/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 694765 lead, implanted: (b)(6) 2008, product id: 383059 lead, implanted: (b)(6) 2008.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was symptomatic when the cardiac resynchronization therapy defibrillator (crt-d) feature which monitors left ventricular pacing amplitude threshold and adjusts outputs would conduct testing.The patient would experience palpitations and extracardiac stimulation.The device feature was turned off and the crt-d remains in use.The patient is a participant in a clinical study.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11759853
MDR Text Key248421408
Report Number3004209178-2021-07047
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169929296
UDI-Public00643169929296
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Reporter Occupation Physician
Type of Report Initial
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/14/2019
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/12/2021
Initial Date FDA Received05/03/2021
Date Device Manufactured08/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6996SQ58 LEAD, 4968-60 LEAD
Patient Outcome(s) Required Intervention;
Patient Age35 YR
-
-