SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Catalog Number 500101
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Hemorrhage/Bleeding (1888); Multiple Organ Failure (3261); Renal Impairment (4499)
Event Type  Death  
Manufacturer Narrative
The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.Investigation summary: intermacs patient registry data collected from july 1, 2019 through september 30, 2020 regarding adverse events was reviewed.Patients were de-identified therefore a one-to-one correlation could not be made between the patient and the lot number of the implanted tah-t.With a review of the available information, there is no evidence of a device malfunction or performance issues that would impact the reported events.Possible clinical factors that may have contributed to these events include the patients pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulants, antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to these events.This device is used for treatment, not diagnosis.(b)(4).(b)(6).
 
Event Description
While supported by the tah-t, the patient experienced the following adverse events as defined by intermacs: 0 days post implant - bleeding.1 days post implant - infection i location pulmonary i type: bacterial.1 days post implant - renal dysfunction.The patient subsequently expired with a cause of death listed as multisystem organ failure (msof) after 6 days of tah-t support.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
don webber
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key11760974
MDR Text Key248523904
Report Number3003761017-2021-00052
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number500101
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/15/2021
Initial Date FDA Received05/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Weight108
-
-