It was reported that the coil needed to be snared to remove.A 2/3mmx2.3cm interlock was selected for use.The target lesion was located in a moderately tortuous vessel.During the procedure, the coil was fed into the microcatheter.The physician was not happy with the catheter position, so the coil was pulled out without detaching.The guidewire was placed into the catheter to reposition, but the coil was unexpectedly in microcatheter and was then deployed in suboptimal position.The coil needed to be snared to remove.The procedure was completed with another of the same device.No further patient complications were reported.
|