MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VLP FOOT IMPLANT TRAY LID; TRAY, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VLP FOOT IMPLANT TRAY LID; TRAY, SURGICAL, INSTRUMENT Back to Search Results
Model Number 71173860
Device Problems Crack (1135); Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2021
Event Type  malfunction  
Event Description
It was reported that the vlp foot implant tray lid plastic coating on the tray lid was cracked and peeling.No case involved; therefore, there was no patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The coating is peeling on the lid, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VLP FOOT IMPLANT TRAY LID
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11770683
MDR Text Key248843486
Report Number1020279-2021-03705
Device Sequence Number1
Product Code FSM
UDI-Device Identifier03596010630018
UDI-Public03596010630018
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173860
Device Catalogue Number71173860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Date Manufacturer Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-