• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problems Mechanical Problem (1384); Calibration Problem (2890)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the mayfield modified skull clamp (a1059) needs calibration and the torque tension does not seem to secure appropriately.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
Device history record: the dhr shows no abnormalities related to the reported failure.Mayfield modified skull clamp (a1059) was returned for evaluation.The reported complaint was confirmed from the evaluation.The set screw in the swivel base was tight and the unit needs to be machined to have heli-coils added to large starburst threads.The evaluation showed that the lock having both rotational and lateral movement and have a residue buildup.The unit needs repair, and replacement of worn/damaged parts.General maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
MDR Report Key11772069
MDR Text Key249034158
Report Number3004608878-2021-00333
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2021
Initial Date Manufacturer Received 04/08/2021
Initial Date FDA Received05/04/2021
Supplement Dates Manufacturer Received05/25/2021
Supplement Dates FDA Received06/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-