It was reported that the svr, core temp and cardiac index values were inaccurate on both (b)(6) 2021.On (b)(6), the ci was 1.0 and the core temp was 33.5.On (b)(6) , the ci was 1.27 and the core temp was 35.The patient was moved to the icu and re-warmed but was getting a temp of 37 with the svr over 5000.The mixed venous blood draws on the patient did not correlate with the readings.However, the pa pressures and waveforms appeared normal.Trouble shooting included changing of the cables, but the inaccurate values persisted.The patient was ultimately placed on a dobutamine infusion.No patient complications were reported.Patient demographics were unable to be obtained at this time.
|
One 831f75 with a monoject limited volume syringe, non-edwards contamination shield and a non-edwards manifold was returned for examination.The reported event of temperature issue was not confirmed.The catheter was submerged in a 37.0c water bath and read 37.0c on a vigilance ii monitor.The thermistor temperature reading accuracy is +/- 0.3c per the vigilance ii manual.The thermistor circuit was continuous.There were no open or intermittent conditions.The balloon inflated clear and concentric and remained inflated for 5 minutes without leakage.All through lumens were patent without leakage or occlusion.The catheter passed the pressure test with lab disposable pressure transducer.A cut down was performed on thermistor connector and there was no visible abnormality.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.It is not known if some procedural factors may have contributed to the event.However, an investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications are well described in the literature.With any hemodynamic monitoring, pressure readings can change quickly and dramatically.Pressure tubing that is used with swan ganz catheters can also be a contributing factor to inaccurate values.In regards to the pressure tubing used with swan ganz catheters, it should be noted that poor dynamic response can be caused by air bubbles, clotting, excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.Pressure readings should correlate with the patients clinical manifestations.It is unknown whether user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|