It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead exhibited variable pace impedance measurements that sometimes rose to high out-of-range pace impedance measurements greater than 2,000 ohms.It was noted that the vectors were previously changed away from the ring as the patient reported feeling stimulation.The field representative was directed to check impedances measurements while the patient was moving.When the patient was leaning forward, impedance measurements went out-of-range.Thresholds were high, however it was noted that the patient had an irregular heart rate and was in atrial fibrillation (af).The field representative was then directed to changing the vector from tip to right ventricular (rv); thresholds were 0.6v@ 1ms, patient did not feel stimulation, and impedance measurements were consistently in the 500 ohm range.Impedances stayed in range when the patient moved to lean forward.This crt-d and lv lead remain in service.No adverse patient effects were reported.
|