This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g3, g6, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined there was rust around the pins on the swivel, the hose that came in with the device connected well with the swivel, the tech could not duplicate the complaint, the swivel and hose were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
|