MEDTRONIC MEXICO S. DE R.L. DE CV BROCKENBROUGH NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a procedure, the transseptal needle did not progress easily.The dilator and needle were removed.It was noted the needle had transfixed the dilator.Two additional dilators were used and difficulty was noted in both progressing the needle through the dilators.When the needle was passed through the distal tip of the dilator, plastic remnants were noted.A different transseptal needle was used and the issue resolved.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: photos of a dilator and transseptal needle were returned or analysis.The photos confirmed that the dilator was punctured by the needle along with plastic dilator pieces.In conclusion, the physical product was not returned for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
MDR Report Key11793815
MDR Text Key251420819
Report Number9617601-2021-00088
Device Sequence Number1
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/06/2021
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number218234907
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/29/2021
Initial Date FDA Received05/07/2021
Supplement Dates Manufacturer Received09/06/2021
Supplement Dates FDA Received10/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-