Model Number RSINT22518X |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problems
Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/04/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
During a procedure one resolute onyx coronary drug eluting stent and one resolute integrity coronary drug eluting stent were attempted to be used to treat a moderately tortuous, severely calcified lesion with 90% stenosis in the distal right coronary artery (rca).The devices were inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The devices did pass through a previously deployed stent.Resistance was encountered when advancing the devices but excessive force was not used during delivery.It was stated that after successfully implanting a resolute onyx stent (ronyx25030x) the physician wanted to stent the distal rca.It was reported that the resolute onyx device failed to cross the lesion, despite the use of scoring balloons.The physician then changed to a resolute integrity device but this also failed to cross the lesion and got caught on the previously implanted stent.The resolute integrity stent was unable to be retrieved and was crushed to the side.No patient injury reported.
|
|
Manufacturer Narrative
|
Additional information: annex d, annex g code added medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: the previously implanted resolute onyx device was not damaged when the resolute integrity device got caught on it.It was stated that the patient underwent cabg.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
This report is being submitted as part of a retrospective review and remediation for capa 564121 per (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|