Catalog Number D2C4711K |
Device Problems
Material Deformation (2976); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The event occurred on an unspecified day of (b)(6) 2021, further described as "recently".The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume folfusor had clear notches/damages (marks/puncture) in the supply line.This was discovered prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from january 21, 2021 - january 22, 2021.H10: the device was received for evaluation.A visual inspection was performed which observed a segment of the tubing line has been damaged.Indentation marks were observed from the damaged areas on the tubing line.The reported condition was verified.The cause of the condition is manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|