• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA-006
Device Problems Migration or Expulsion of Device (1395); Off-Label Use (1494)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2021
Event Type  Injury  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2021, a 6mm amplatzer duct occluder as selected for implant.Post implant it was reported that the device embolized from a paravalular leak (pvl) due to an unknown mechanical valve.The device was snared successfully using a 12f trevisio and a ensare gooseneck snare.A new 9mm amplatzer duct occluder was successfully implanted.The patient was reported to be in stable condition.
 
Manufacturer Narrative
An event of device embolization was reported.The investigation confirmed the device met functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Type of Device
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key11805301
MDR Text Key249986948
Report Number2135147-2021-00185
Device Sequence Number1
Product Code MAE
UDI-Device Identifier00811806011134
UDI-Public00811806011134
Combination Product (y/n)N
PMA/PMN Number
P020024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9-PDA-006
Device Catalogue Number9-PDA-006
Device Lot Number7662619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2021
Was the Report Sent to FDA? No
Date Manufacturer Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight70
-
-