Model Number 9-PDA-006 |
Device Problems
Migration or Expulsion of Device (1395); Off-Label Use (1494)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2021, a 6mm amplatzer duct occluder as selected for implant.Post implant it was reported that the device embolized from a paravalular leak (pvl) due to an unknown mechanical valve.The device was snared successfully using a 12f trevisio and a ensare gooseneck snare.A new 9mm amplatzer duct occluder was successfully implanted.The patient was reported to be in stable condition.
|
|
Manufacturer Narrative
|
An event of device embolization was reported.The investigation confirmed the device met functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|
|