BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, UNIV ENDS,; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number H93821 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter phone no.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a fluid transfer tube set was leaking with unspecified medicine.The leak was coming from the white area of the set that fits into the pump.A new set was replaced and the issue resolved.This issue was identified during priming.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H4: device manufactured between july 06, 2020 to july 08, 2020.H10: the device was received for evaluation.Unaided eye visual inspection was done which observed a hole at a torn section of the silicone tubing of the tube set.A functional testing was performed which revealed a leak at the silicone tubing area where the hole was located.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|