Model Number VS101012P |
Device Problems
Detachment of Device or Device Component (2907); Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection of the first image showed a surgical procedure being performed with the image on a monitor of some organs with an unknown plastic piece in view, while the second image showed a slightly different angle view of the image on a monitor of some organs with an unknown plastic piece in view.It was reported that the obturator was damaged, and a component disengaged from the device into the surgical cavity.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, laparoscopic inguinal hernia repair, a piece of the top plastic part of the trocar sleeve broke off and landed in the abdomen of the patient.The surgeons found the little piece during the last optical check around inside the patient prior to pulling out trocars and closing the port holes.The trocar piece was retrieved using laparoscopic graspers.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection of the first image showed a surgical procedure being performed with the image on a monitor of some organs with an unknown plastic piece in view, while the second image showed a slightly different angle view of the image on a monitor of some organs with an unknown plastic piece in view.It was reported that the obturator was damaged, and a component disengaged from the device into the surgical cavity.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|