SMITH & NEPHEW, INC. ACCU-PASS SUTURE SHUTTLE 70 DEGREE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
Model Number 72200419 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during surgery, the "accupass suture shuttle 70°" did not pass the suture properly.The procedure was successfully completed without delay using a back-up device.No patient injuries or other complications were reported.
|
|
Manufacturer Narrative
|
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Internal complaint reference: case-(b)(4).
|
|
Manufacturer Narrative
|
H10: h3, h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A visual inspection revealed the device was returned outside of original packaging.The monofilament was received halfway inside of the handheld instrument.No damage visible.A functional evaluation revealed the device passes suture as intended.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|