Device evaluated by mfr.: returned product consisted of an angiojet solent omni catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The shaft showed no damage.Functional testing was completed per device preparation.The pump was inserted into the ultra drive unit console.The pump and device primed and ran as designed for 120 seconds in the thrombectomy mode.The devices pressure was within the normal range.The device functioned as design with no abnormalities noticed.An aspiration test was performed by inserting the devices tip into a 100cc beaker.The device was activated, and the device removed fluid as designed.No leaks were noticed during functional analysis.Inspection of the remainder of the device damage, revealed no damage or irregularities.
|