MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVPROPLUS-29US
Device Problems Perivalvular Leak (1457); Material Split, Cut or Torn (4008)
Patient Problems Low Blood Pressure/ Hypotension (1914); Cusp Tear (2656); Valvular Insufficiency/ Regurgitation (4449); Aortic Valve Insufficiency/ Regurgitation (4450)
Event Date 05/12/2021
Event Type  Injury  
Manufacturer Narrative
Product analysis: the valve remains implanted therefore no product analysis can be performed.Conclusion: without return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, moderate paravalvular leak (pvl) was noted.A balloon aortic valvuloplasty (bav) was performed with a 25 mm balloon and moderate pvl and aortic insufficiency was still noted.The patient remained stable.A bav was performed with a 26 mm balloon, causing the leaflets to tear.A second valve was implanted.Hemodynamics remained stable.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Additional information was received that following the 25 mm balloon aortic valvuloplasty (bav), mild to moderate central aortic insufficiency was noted.Following the 26 mm bav, severe central aortic insufficiency was noted.One inflation was performed for an unknown amount of time using a syringe.The inflation pressure was unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information was received that a 60 cc syringe was used to inflate the balloons, however the inflation pressure was not able to be determined behind the balloon.It was reported that the physician was not aggressively filling the balloon.The patient's diastolic pressure dropped following the bav, along with the severe central aortic insufficiency, which lead the team to suspect the leaflet tear, however the leaflet tear was not confirmed.Updated b.5 - description of the event.Updated h.6 - patient codes.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information was received which indicated that the physician attempted to keep the balloon position low.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Paravalvular leak can be caused by a variety of factors, including valve positioning, patient anatomy, or presence of pre-existing patient conditions, and a conclusive cause could not be determined from the limited information available.Hypotension is a known potential adverse effect per the ifu.It is an effect that is highly dependent on the patient's pre-procedural condition and can occur despite a normally-functioning device or model implant procedure.However, based on the event description information, preliminary assessments of the cause(s) of the leaflet damage reported in the event description, suggest that the post balloon aortic valvuloplasty (bav) intervention is possibly among the factors leading to the damage of the leaflet, as it was stated that the user used both a z-med balloon, a compliant / semi compliant balloon and a true balloon, which is a non compliant balloon.Per instructions for use (ifu) table 3: post-implant balloon dilatation sizing, the maximum balloon size chosen for dilatation using a compliant /semi-compliant balloon for the 29mm evolut pro+ valve, is 23-26mm with an applied inflation pressure of no greater than 2 atm.And the maximum balloon size for the non compliant balloon (true balloon), is 24-24mm.So as the user reported, most likely the leaflet damage occurred during use with the 26mm true balloon as it is not listed as a recommended size per the ifu.In addition, it is unknown what the inflation pressure was during use with either balloon.Per the ifu, the inflation pressure should not exceed 2 arm, and the balloon should not exceed 1mm more than the tav waist diameter.As overexpansion of the narrowest portion (waist) of the evolut pro+ tav beyond the levels set forth in table 3 of the ifu has been demonstrated through bench data to cause damage to the bioprosthetic leaflets.These complaints of damage to the leaflets during post implant balloon dilatation have been reported in some clinical cases resulting in moderate to severe aortic insufficiency.Update data: method code, results code and conclusion code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Corrected: h9.The previous correction number was submitted in error.The number should be 2025587-10-28-2020-001-c.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO PLUS VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11836251
MDR Text Key251215406
Report Number2025587-2021-01558
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000211127
UDI-Public00763000211127
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/08/2022
Device Model NumberEVPROPLUS-29US
Device Catalogue NumberEVPROPLUS-29US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2021
Initial Date FDA Received05/17/2021
Supplement Dates Manufacturer Received05/20/2021
06/01/2021
07/02/2021
07/02/2021
03/20/2022
02/23/2023
Supplement Dates FDA Received05/25/2021
06/07/2021
07/06/2021
07/14/2021
10/28/2022
02/24/2023
Date Device Manufactured10/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2025587-10-28-2020-001-C
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexMale
-
-