MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX-FX05RE
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: not required for product code.Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: clinical engineer.Pma/510(k): k130280.The actual sample was received for evaluation.Visual inspection revealed no break or no other anomaly in the appearance.Air was blown from the gas inlet of the actual sample.As a result, liquid was observed adhering to the housing on the gas outflow side; however, as it was a very small amount, it was not possible to identify the liquid.The actual sample, after having been rinsed and dried, was tested for its oxygenation and co2 removal performance in accordance with the factory's inspection protocol.No anomaly was observed, and the obtained values met the control criteria.Bovine blood conditions: hb:12g/dl, temp.: 37°c., ph:7.4, svo2:65%, pvco2: 45mmhg.Circulation conditions: blood flow rate: 2l/min and 1l/min, v/q=1, fio2=100%.O2 transfer volume: @2l/min= 124 ml/min.@1l/min= 70 ml/min.Co2 removal volume: @2l/min= 82 ml/min.@1l/min= 53 ml/min.A review of the device history record and product-release judgement record of the involved product code/lot# combination was conducted with no findings.The pump record review revealed: the blood flow rate was found to be constant at about 0.9-1.0 l/min.In addition, the arterial and venous pressures were found following the changes in blood flow rate and not change drastically.Therefore, it was unlikely that blood clot formation was a factor.When pao2 decreased, fio2 and gas flow rate were increased.Pao2 increased temporarily, and then decreased gradually.Therefore, it was likely that the gas exchange performance decreased for some reason.It could not be clarified if the degradation in gas transfer performance was relevant to the event noted in the pump record (start of rewarming, ao declamp) or the administration of heparin/ffp.The pao2 was found markedly fluctuating from around 16:00 (at the beginning of the provided pump record).The blood gas data after 16:00 showed a gradual decrease in svo2, suggesting that this may have contributed to the decrease in pao2.Ifu states: a phenomenon called wet lung may occur when water condensation occurs inside fibers of microporous membrane oxygenators with blood flowing exterior to the fibers.This may occur when oxygenators are used for a longer period of time.If water condensation and/or a decrease in pao2 and/or an increase in paco2 is noted during extended oxygenator use, briefly increasing the gas flow rate may improve the performance.Increase gas flow rate, to 5 l/min for 10 seconds.Do not repeat this flushing technique, even if oxygenator performance is not improved.Measure blood gases and make necessary adjustments as follows.A.Control pao2 by changing concentration of oxygen in ventilating gas using gas blender.To decrease pao2, decrease fio2.To increase pao2, increase fio2.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.The investigation results verified the returned sample was of normal product.It is likely that from the liquid flowed out from the gas channel, it was presumed that hydrophilization of the fiber occurred disrupting the balance maintained in the micropores of the fiber between the blood surface tension and the gas pressure, resulting in plasma leak and decrease in the gas exchange performance.The decrease in the gas exchange performance of the actual sample caused a decrease in the oxygen supply to the patient, resulting in the decrease in svo2 and pao2.However, the exact cause of the reported event cannot be definitively determined based on the available information.(b)(4).
 
Event Description
The user facility reported that four hours after the start of circulation, oxygenation performance dropped sharply.Increasing gas flaw up to 10l/min could not improve the situation.Gas flushing was repeated three times, although that could not either.Replacement of the oxygenator was discussed; however, since it was already at the end of the case, it was judged that there would be no problem, and the oxygenator in question was continued to be used as it was.The procedure was finished safely and was completed successfully.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key11847147
MDR Text Key265198944
Report Number9681834-2021-00080
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue NumberCX-FX05RE
Device Lot Number201109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-