ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-01514.
 
Event Description
It was reported that threaded tip on inserter broke off in the 32mm impactor when surgeon was impacting liner.All parts were retrieved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6.   one g7 str monoblock shell insrtr was returned and evaluated.Upon visual inspection the ball impactor has a square indentation on the od of the device and there are fractured threads inside the threaded portion of the device.The returned inserter has impact marks on the strike plate and some scuffing on the shaft.The threaded tip of the device had fractured from the device.A review of the device history records identified deviations or anomalies during manufacturing, the deviations or anomalies would not have attributed to the event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11847716
MDR Text Key251379566
Report Number0001825034-2021-01513
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304532694
UDI-Public(01)00880304532694(11)150911(10)252270
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number252270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/04/2021
Initial Date FDA Received05/19/2021
Supplement Dates Manufacturer Received09/07/2021
Supplement Dates FDA Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-