• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40C
Device Problems Intermittent Capture (1080); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but was not yet available.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that crosstalk was observed on the implantable cardiac defibrillator.
 
Event Description
New information received notes the patient was having 2-3 second pauses on telemetry due to pacemaker inhibition while in hospital following knee surgery.Programming changes to sensitivity were made and to prevent pauses and inhibition.The physician was to continue monitoring the patient.The patient did not state they were having any symptoms as a result of the observation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11860292
MDR Text Key251843302
Report Number2017865-2021-18693
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508261
UDI-Public05414734508261
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2018
Device Model NumberCD3365-40C
Device Catalogue NumberCD3365-40C
Device Lot NumberA000016785
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/01/2021
Initial Date FDA Received05/20/2021
Supplement Dates Manufacturer Received05/24/2021
Supplement Dates FDA Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age85 YR
-
-