Manufacturer Narrative
The navio handpiece intended for treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation could not be performed due to broken/damaged parts.The gears would not turn.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is handpiece motor prematurely failing after autoclaving.Further investigation into the reported failure is being conducting to determine if additional actions are required.Internal complaint reference number: (b)(4).
 
Event Description
It was reported that before a navio procedure, they ran a handpiece test and it could not be completed.The max torque value was 98, and the t2 iterations were 0/10.They replaced the handpiece for the case, so there was no impact to the case.After the case, they inspected the handpiece.Although nothing seemed to be broken, they could not actuate the gears manually.Even after applying a couple of drops of lubricant, the gears could not be actuated.No other complications were reported.The investigation found that the gears were broken/damaged and would not turn, and motor prematurely failing after autoclaving.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11860479
MDR Text Key251951299
Report Number3010266064-2021-00398
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-