MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. UNK_CARTRIDGE_PLATINUM_UNKNOWN; UNFOLDER PLATINUM 1 SERIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. UNK_CARTRIDGE_PLATINUM_UNKNOWN; UNFOLDER PLATINUM 1 SERIES Back to Search Results
Model Number 1MTEC30
Device Problems Crack (1135); Difficult to Advance (2920); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2021
Event Type  malfunction  
Manufacturer Narrative
If implanted, give date: not applicable as this is not an implantable device.  if explanted, give date: not applicable as this is not an implantable device.  the device is not available to be returned for analysis as the device was discarded.Therefore, no further investigation can be performed.If there is any further relevant information received, a supplemental medwatch will be filed.Device manufacture date: unknown as lot number was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the intraocular lens (iol) advanced only halfway and did not advance any further when doctor began to advance the iol into patient's right eye.It was observed that the plunger had penetrated the cartridge under the iol.Additional information received confirmed that the plunger went under the iol in the cartridge after it had inserted the iol halfway into the patient's eye and the plunger had penetrated the bottom of the cartridge.The cartridge had a hole from the plunger on the bottom about 4mm from the tip.The lens was 50% in the eye and 50% out of the eye including the cartridge tip and would not advance any further.The incision was enlarged and a new lens of same model and diopter was implanted without any issues.There was no injury and no other medical/ surgical intervention was required.The procedure had a delay of 10 minutes and the patient had fully recovered.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK_CARTRIDGE_PLATINUM_UNKNOWN
Type of Device
UNFOLDER PLATINUM 1 SERIES
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
road 402 north, anasco ind. pk
anasco, PR 00610
7142478552
MDR Report Key11868043
MDR Text Key264141808
Report Number2648035-2021-07775
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(10)UNKNOWN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/23/2021
Initial Date FDA Received05/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-